The microscope is composed of several sub-assemblies including:
- The power control unit houses the power switch,
knobs to control light intensity and the speed of the X-Y, Zoom and
Focus controls, buttons to move the column up and down, and buttons
to center the X-Y and Focus.
- The two foot switches control zoom, focus,
and X-Y movement.
- The two hand controls also control zoom,
focus, and X-Y movement.
- The heavy duty H-base houses the casters and
stabilizing locks.
- The Vertical Column moves up or down.
- The Horizontal Arm assembly houses the power control unit and fuses.
- The Swivel Arm assembly holds
the fiber optics.
- The fiber optic lighting provides optimal
illumination.
- The X-Y Translation Arm gives 4"
x 12" of planar movement.
- The Head Assembly houses the two
microscope binocular bodies.
- The video and camera systems allow for documentation
and easy viewing by an audience.
Return to Index
POWER CONTROL PANEL

- X-Y Speed Control Knob
- X-Y Center Button
- X-Y Orientation Selector
- Zoom Speed Control Knob
- Focus Center Button
- Focus Speed Control Knob
- Column Up and Down Buttons
- Power Switch with Indicator Light
- Light Intensity Control Knob
- The controls on the top row are for the X-Y Translation Arm. This arm moves the microscope
eyepieces in a plane over the work area. The X-Y Speed knob is
at the top left corner of the power control panel. It controls the responsiveness
of the foot switch joystick and the hand control yellow pushbuttons.
"0" is the slowest setting used for very fine adjustments.
"5" is the fastest setting for quick response. This is a continuous
control and can be set anywhere, not just on the numbered positions.
- Pushing the X-Y Center green button
at the top center of the panel will move the scope heads to the centered
position to reset the reference point and have the maximum range of
motion. The button lights when the travel is centered.
- The X-Y Orientation Selector switch
at the top right assures that X-Y control inputs deliver reasonable
head movements at various head positions.
- If the blue Praeger handle on the head carrier is pointed to the
"L Selector" position on the X-Y Translation Arm, set
the X-Y Orientation Selector switch to "L." Then the hand
controls for A and B-sides operate intuitively: left moves left,
right moves right, up moves away and down moves toward.
- If the blue Praeger handle is pointed toward the "R Selector"
position on the X-Y translation arm and the X-Y Orientation Selector
is set to "R," the hand and foot switches still function
intuitively.
- When the blue handle is set to "N Selector," set the
X-Y Orientation Selector to "N" for proper functioning.
- If the X-Y Orientation Selector setting is different than the
position of the Praeger handle, the controls will behave weirdly
- pushing the left button moves the microscope right or away.
- The Zoom Speed knob is on the left side
of the center row. This potentiometer sets the responsiveness of the
Zoom controls on the foot and/or hand switches. "0" is the
slowest setting used for very fine adjustments, and "5" is
the fastest setting for quick response. This is a continuous control
and can be set anywhere, not just on the numbered positions.
- The Focus controls are on the middle row of
the power control panel. Pushing the green Focus Center button
moves the focus body to the center of its travel. The button lights
when the focus is in the centered position.
- The Focus Speed control knob sets the
responsiveness of the Focus controls on the foot and/or hand switches.
"0" is the slowest setting used for very fine adjustments.
"5" is the fastest setting for quick response. This is a continuous
control and can be set anywhere, not just on the numbered positions.
- The yellow Column Up and Down buttons
move the column up and down to provide a comfortable working setup.
Column height is adjusted with the Column toggle switch on the Carrier Assembly.
- The Power Switch is in the center of
the bottom row. Up is ON (green light comes on); down is OFF. If the
light does not come on, check the power cord and fuses.
- The Light Intensity control knob is
at the lower right of the panel and controls the intensity of the fiber
optic light. "0" is the dimmest setting, and "5"
is the brightest. This is a continuous control and can be set anywhere,
not just on the numbered positions.
The fuses are in the back of this unit and are covered in the Fuse Legend section.
Return to Index
FOOT SWITCHES

The foot switches control most of the motorized functions of the microscope
to free the surgeon's hands.
- The joystick at the top of the foot switch moves the X-Y Translation
Arm left, right, away and toward.
- A raised bar provides a footrest and allows heel-or-toe rocking
action for zoom and focus.
- The toe controls Focus: Pressing the left side focuses up,
the right side focuses down. The Focus Speed knob on the control panel
sets the sensitivity.
- The heel controls Zoom: Pressing the left side zooms in, the
right side zooms out. The X-Y Speed knob on the control panel sets the
sensitivity.
The Z880 microscope has two joystick foot controls with the same functionality
for A and B-side.
The cables for the foot switches may be disconnected from the Horizontal
Arm. Always move the foot switches by the handles provided, not by
the foot switch cable. Use the hangers on the column for handy storage.
Return to Index
HAND SWITCHES

The Z880 has easy-to-use hand controls for timely and accurate set-up.
When the microscope is turned on, all buttons on the hand switches are
lit.
- The top buttons control focus: left
up, right down. The Focus Speed knob
on the control panel sets sensitivity.
- The middle buttons control zoom: left
in, right out. The Zoom Speed knob
on the control panel sets sensitivity.
- The four yellow directional buttons control the travel of the X-Y
Arm. The left button moves the head left,
the right button right, the top
button away and the bottom button towards. Set
the X-Y Orientation Selector to the same position as the Praeger handle
or else the directions will be confused.
If desired, loosen the knurled nut at the back of the hand switch to
remove for placement of a sterile covering. For proper operation of the
microscope, tighten the nut to clamp the hand switches securely in the
holder.
Return to Index
H-BASE

The heavy-duty H-base houses the caster mechanism and stabilizing locks.
Step on both black buttons to lock the casters. Unlock by stepping on
the raised edge of the silver plates. Keep the casters locked unless moving
the microscope. Please refer to the Storage Instructions section for information on safely moving the microscope.
Return to Index
VERTICAL COLUMN
The motorized column allows 20" of fingertip-controlled travel,
making it easy to achieve gross focus. The yellow
buttons on the power control panel move the column up and down. The toggle switch on the Carrier Arm also controls column movement
and has a blue sterilizible cover.
Two hangers provide convenient storage for
the foot switches when not in use.
Return to Index
HORIZONTAL & SWIVEL ARMS

The Swivel Arm directly supports the X-Y and Head Assembly and houses
the fiber optic illumination source. The Horizontal Arm houses power control
unit. Both arms move freely to easily position the heads over the work
area. Four black knobs allow you to adjust and lock the positioning of
the microscope. Turn counter-clockwise to loosen the brake and allow free
movement, turn clockwise to tighten the brake.
- Looking at the power control panel,
the Horizontal Arm brake is the black knob on the left side.
The Horizontal Arm swivels 330° about the Vertical Column.
- The Swivel Arm brake is the black
knob located on the Horizontal Arm directly underneath the fiber optic
light source. The Swivel Arm rotates 390° about a pivot point in
line with the light source.
- The X-Y Rotation brake is the black
knob located at the end of the Swivel Arm, directly above the X-Y. The
X-Y Translation Arm pivots about its center.
- The Carrier Rotation brake is the
black knob on the Carrier Arm Coupling at the top of the Carrier Assembly,
just above the Column Toggle switch. While the Carrier Arm can be set
in any position, the Praeger handle should point to the "L,"
"N," or "R" position (with the X-Y Orientation Selector
on the power control panel set the same) for proper use of the X-Y controls.
Return to Index
ILLUMINATION

The fiber optic system delivers optimal illumination, decreases reflections
and results in a bright shadow-free field.
On top of the Swivel Arm assembly is the remote halogen light source
with a plug-in lamp module. Remove this lamp module by pulling it out
by the handle. To replace, slide module into position with the handle.
Do not touch the halogen bulb since skin oils can cause the bulb to fail
sooner.
To change the halogen bulb, see the service notes.
Twin fiber optic bundles pass light between the binocular bodies to illuminate
the work area. Take care to not bend or stretch the bundles which, although
flexible, contain tiny glass fibers which could easily break, reducing
the intensity of the light. If the fiber optic bundles have to be removed
from the light source unit, pull them out by grasping the metal sleeve
on the end of the bundle rather than the bundle itself.
The Light Intensity knob in the lower right corner
of the power control panel sets the brightness of the light.
Return to Index
X-Y TRANSLATION ARM

Unlike pendulum movement, which tilts the microscope and can affect the
focus, Medlink USA microscopes have true X-Y single plane movement.
Foot and hand switches position the microscope heads within the 4"
x 12" range. To move the binocular heads:
- Loosen the Carrier Rotation brake by
turning counter-clockwise.
- Point the blue Praeger handle to the "L," "N,"
or "R" position. With the short axis pointing away from you,
"L" is with the Praeger handle to the left, "N"
is to the back, and "R" is to the right.
- Turn the Carrier Rotation brake clockwise to tighten.
- Set the X-Y Orientation Selector on the power
control panel to the same setting as the Praeger handle.
- Adjust the speed of the X-Y Translation Arm between "0"
(slowest) and "5" (fastest) with the X-Y
Speed knob on the power control panel.
- Push the X-Y Center button (top green button)
on the power control unit to center the X-Y within its field of travel.
The button lights when the X-Y Translation Arm is centered.
- Move the microscope heads with the joystick foot switches or the hand
switches. The functioning is intuitive. Left moves left, right moves
right, up moves away and down moves towards. If the behavior is weird,
check that the X-Y Orientation Selector and the position of the Praeger
handle agree.
Return to Index
CARRIER ASSEMBLY

Horizontal and vertical tilt provide the surgeon with the unlimited flexibility
to use at any angle needed for executing the most complicated modern procedures.
There are three parts to the Carrier Assembly: the Carrier Arm Coupling
(which has the Carrier Rotation brake), the Carrier Arm, and the Focus
Body.
- The Carrier Rotation brake is the black knob on the Carrier
Arm Coupling at the top of the Carrier assembly, just above the Column
Toggle switch. This allows rotation of the carrier. While the Carrier
Arm can be set in any position, the Praeger handle should point to the
"L," "N," or "R" position (with the X-Y
Orientation Selector on the power control panel set the same) for proper
use of the X-Y controls.
- The Column Control switch the top
part of the Carrier Arm raises or lowers the vertical height/coarse
focus. It has a sterilizible blue cover. The yellow Column Up and Down
buttons on the power control panel also set column height.
- The blue Praeger handle located on the back of the Carrier
Arm is used for gross rotation of the head and carrier on a X-Y plane.
The Praeger handle can be unscrewed and removed for sterilization. Line
up the handle with the left (L), back (N) or right (R) side of the X-Y
Translation Arm and set the X-Y Orientation Selector on the power control
panel to the same setting to allow transparent use of the hand or foot
controls.
- Dual hand controls mounted on
the Carrier Arm just below the Praeger handle control focus, zoom and
X-Y movement.
- The Vertical Tilt knob with blue sterilizible cover is directly
underneath the Praeger handle and controls 60° of inclination.
- The 120° Horizontal Tilt knobs are located at the base
of the vertical Carrier Arm. The knob is to the left of the A-head,
to the right of the B-head. The blue covers are sterilizible.
Return to Index
TWIN HEAD ASSEMBLY

Dual zoom and independently-controlled assistant fine focus allow both
users to independently choose the magnification for a wider, all inclusive
view. This feature gives both the surgeon and assistant much more versatility,
enhancing the quality of teamwork. The Z880 has Inclining Binoculars with
manual Pupillary Distance (P.D.) adjustment and B-Side Fine Focus.
- Independent Inclining Binoculars allow a 90° range of motion
to position the eyepieces vertically, horizontally, or anywhere in between.
To move the binoculars, gently but firmly twist the eyepieces to the
desired location.
- Head A and B each have 140° independent rotation around
their own optical axis for comfortable viewing in various surgical angles
including opposite, side by side, 90° apart, or any other convenient
point. Rotate the A-side and the B-side binoculars for a comfortable
working position.
- Set the rubber eyecups on the high eyepoint eyepieces down
for use with glasses and up for use with out glasses. To change the
position, carefully peel up the rubber or fold it back on itself.
- The Diopter Control is on each eyepiece
just below the rubber eyecups. If you wear eyeglasses or do not need
vision correction, set the Diopter Controls to "0" on each
eyepiece. For vision corrections up to +/- 4 diopters, dial in the setting
for each eye. Turn to the right for positive adjustments, left for negative
adjustments.
- The knurled ring below the Diopter Control is for removing the eyepieces.
- The Pupillary Distance (P.D.) control knob
sets the distance between the eyepieces. The knob is located to the
right of the eyepieces and has a sterilizible blue cover. The scale
located between the eyepieces reads in millimeters. Set the P.D. to
match your prescription or twist the knob for comfortable viewing.
- At the base of each binocular head is a black ring that shows the Zoom (magnification) range. The current magnification level is
indicated by the white mark on the top ring. Do not twist the ring.
To zoom in or out, press the Zoom pedal on the foot switch or press
the green zoom buttons on the hand controls. The Zoom Speed knob on
the power control panel sets the responsiveness of the Zoom controls
for both A-side and B-side. The A-side controls affect only the A-side
zoom, the B-side controls affect only the B-side zoom. The magnification
range from 4-20 assumes the microscope has 10x eyepieces. Optional 15x and 20x eyepieces are available.
- The twin binocular assembly has a three inch fine focus track on the base of the carrier. Pushing the green Focus Center button on
the power control panel moves the assembly to the center of the track
to center the focus. Use the fine focus pedal on the foot switch and
the green focus buttons on the hand controls to focus the microscope.
The Focus Speed knob on the power control panel sets the responsiveness
of the focus controls. Since both binocular scopes are parfocal, only
one surgeon should control electronic focus.
- When one microscope is in focus, the other should also be in focus.
To account for minor differences, use the B-Side
Fine Focus Adjustment which is controlled by the two opposing black
knobs (with sterilizible blue covers) at the bottom of the B-Side head.
When the white marks line up, the focus is centered. It can be adjusted
up or down approximately 1/2 inch.
Return to Index
FOCUS PROCEDURE

Both surgeons can zoom in and out independently without having to refocus.
Work first with Head A:
- Set the Diopter Control on each eyepiece to
"0."
- Zoom in to the highest magnification.
- Focus the microscope until the image is sharp.
- Zoom out to the lowest magnification.
- Adjust the Diopter Control on each eyepiece individually to attain
the sharpest image.
NOTE: If the user has to wear glasses to correct an astigmatism, it is
suggested that they operate the microscope while wearing their glasses
and using the high point eyepieces. Return to the highest point of magnification
and refocus microscope if necessary.
Use the manual B-Side fine focus to manually fine
tune the focus with the two opposing knobs at the bottom of the B-Side
head.
Return to Index
VIDEO AND CAMERA SYSTEMS
The 35mm camera and video systems operate with independent zoom controls.
Since the system requires no beamsplitter, there is no loss of illumination
to either surgeon.
- Install the video camera on the standard c-mount between the
A and B heads.
- Below the c-mount is the iris control. Most video systems have
automatic image controls, but the reflectivity of the body and the extreme
brightness of the fiber optic lighting may be outside of the adjustment
range of the video camera. Use the iris control to adjust the video
image for good picture - not washed out, not too dark.
- The large knurled nut below the iris control is the mounting ring
for the video system.
- The black knob on a shaft is the independent video system zoom
control.