Medlink 800 Series Microscope Fuse Legend

WARNING:

For protection against fire hazards, replace all fuses with new fuses of the exact same type and rating.

Fuses are located under the back cover on the back of the Horizontal Arm.

To replace a fuse, remove the back cover by removing the four Phillips screws and sliding the cover back. Use a small slotted screwdriver to remove the grey fuse holders. Push in, turn counter-clockwise 45° and remove.

To insert a new fuse, put a new fuse in the fuse holder, slide the fuse holder into the panel, and gently rotate counter-clockwise until it seats in further. Use a small slotted screwdriver to push it in, turning clockwise until it stops.

FUSE

#

CONTROLS:

IF OPEN:

REPLACE WITH:

1

P.C. Board (Focus, Zoom, and X-Y Functions) Power light and everything except the fiber optic illumination would fail to operate.

250V UL SA;

MDA 3

2

Column Motor
Column-Up / Column - Down features would fail to operate.

250V UL SA;

MDA 6 1/4

3

Fiber Optic Light Source
Fiber Optic Light Source would fail to operate.

250V UL SA; 

MDA 15

N

Input Voltage to Power Supply
ALL FUNCTIONS and ALL ILLUMINATION would fail to operate.

250V UL SA; 

MDA 4

L

Input Voltage to Power Supply
ALL FUNCTIONS and ALL ILLUMINATION would fail to operate.

250V UL SA; 

MDA 4

Note: MDA is a slow blow ceramic type fuse. MDL type fuses may also be used.

CAUTION:

The voltage selector switch is pre-adjusted at the factory and should never need re-adjustment. Under no circumstances should the selector switch be set to a voltage lower than the incoming line voltage as measured at the receptacle that the microscope is plugged into.

Before making any adjustments to the selector switch, contact Medlink USA's Service Department. Failure to do so could cause serious and costly damage to the unit.