The floor stand is shipped partially assembled. The initial assembly
procedure should be completed by a trained Medlink USA representative.
The following procedures are given here for reference and to assist qualified
individuals to disassemble portions of the microscope.
For the following procedures, refer to System
Parts ID:
To Change Fiber
Optic Bulb:
- Remove slide-in lamp module from housing.
- Take care to not touch with bare hands since skin oil will shorten
bulb life.
- Tip bulb back out of recess, slide out away from leaf springs.
- Pull bulb away from bulb holder.
- Unplug bulb from socket.
- Plug new bulb into socket.
- Put bulb into holder aligning tab with slot in holder.
- Re-install slide-in lamp module into housing.
Return to Index
To Disconnect Foot
switches:
- Place thumb and index finger on ring of connector and pull down gently.
Return to Index
-
CAUTION:
-
To avoid personal injury and damage to the equipment,
it is strongly recommended that the following procedures be completed by
at least two individuals.
To Disconnect and Remove Microscope Head and Carrier Assembly:
- Place thumb and index finger on ring of Head connector, pull up gently.
- Place thumb and index finger on fiber optic ferrule at the end near
fiber optic light housing, pull up gently.
- Loosen tension knob (counter-clockwise).
- Rotate so Carrier is not under X-Y Translation Arm.
- Tighten tension knob (clockwise).
- Loosen locking knob (counter-clockwise) 3 turns. It is the small knob
pointed out 45° to the tension knob.
- Pull back locking lever, lift head up and off by holding the black
hub that holds the column joystick, and the Carrier.
- Put down very carefully on clean moderately hard surface.
Return to Index
To Disconnect and Remove X-Y
Transfer Arm Assembly:
-
NOTE:
-
To remove X-Y Translation Arm, the Microscope
Head and Carrier Assembly MUST BE REMOVED FIRST.
- Place thumb and index finger on ring of X-Y connector and pull up gently.
- Loosen tension knob (counter-clockwise)
- Rotate X-Y Translation Arm until the long axis of the X-Y is pointed
parallel to the Swivel Arm.
- Tighten tension knob (clockwise).
- Loosen locking knob (counter-clockwise) 3 turns. It is the small knob
pointed out 45° to the tension knob.
- Pull back locking lever, lift up and off.
Return to Index
Tools Needed:
- #2 Phillips Head Screwdriver
- 5/16" Hex Head Wrench
- 5/32" Hex Head Wrench
- Needle-Nosed Pliers
To Remove Control
Panel:
- Unplug microscope from wall socket.
- Remove the four Phillips screws (one in each corner) with a #2 Phillips
screwdriver.
- Grasp the Control Panel in one hand firmly, while with the other hand
disconnecting the four brown connectors (12 pin, 14 pin, 14 pin, 20 pin),
one white connector (9 pin), and four wires plugged into the power switch
(2 white, 2 black).
Return to Index
To Remove Power
Supply:
- Unplug microscope from wall socket
- Remove the Control Panel.
- Remove four Phillips screws from the back cover with a #2 Phillips
screwdriver.
- Remove two Phillips screws for the venting cover on top of the horizontal
arm.
- Disconnect one white connector (7 pin) through the venting cover.
- Disconnect one white connector (8 pin) through the back cover.
- Remove one Phillips screw in the side of the horizontal arm with a
#2 Phillips screwdriver, using a pair of needle-nosed pliers to hold the
nut on the inside of the arm.
- Remove four socket head cap screws on the back of the black heatsink
with a 5/32" hex head wrench.
- Remove the grounding screw for the horizontal arm with a #2 Phillips
screwdriver.
- Pull the Power Supply out of the back of the arm gently while being
extremely careful not to damage any of the connectors.
Return to Index |